Oxygenator Impact on Ceftaroline in Extracorporeal Membrane Oxygenation Circuits

Objectives: To determine the oxygenator impact on alterations of ceftaroline in a contemporary neonatal/pediatric (1/4-inch) and adolescent/adult (3/8-inch) extracorporeal membrane oxygenation circuit including the Quadrox-i oxygenator (Maquet, Wayne, NJ). Design: Quarter-inch and 3/8-inch, simulated closed-loop extracorporeal membrane oxygenation circuits were prepared with a Quadrox-i pediatric and Quadrox-i adult oxygenator and blood primed. Additionally, 1/4-inch and 3/8-inch circuits were also prepared without an oxygenator in series. An one-time dose of ceftaroline was administered into the circuits, and serial pre- and postoxygenator concentrations were obtained at 5 minutes, 1-, 2-, 3-, 4-, 5-, 6-, and 24-hour time points. Ceftaroline was also maintained in a glass vial, and samples were taken from the vial at the same time periods for control purposes to assess for spontaneous drug degradation. Setting: A free-standing extracorporeal membrane oxygenation circuit. Patients: None. Intervention: Single dose administration of ceftaroline into closed-loop extracorporeal membrane oxygenation circuits prepared with and without an oxygenator in series with serial preoxygenator, postoxygenator, and reference samples obtained for concentration determination over a 24-hour study period. Measurements and Main Results: For the 1/4-inch circuit with an oxygenator, there was 79.8% drug loss preoxygenator and 82.5% drug loss postoxygenator at 24 hours. There was a statistically significant difference (p