Τετάρτη 1 Φεβρουαρίου 2023

Effect of finish line location and saliva contamination on the accuracy of crown finish line scanning

alexandrossfakianakis shared this article with you from Inoreader

Abstract

Purpose

Intraoral scanners are used widely as an alternative to conventional impressions, but studies on the influence of finish line location and saliva contamination on scanning trueness are lacking. The purpose of this in vitro study was to evaluate the influence of finish line location and saliva contamination on the scanning trueness of crown finish lines.

Materials and Methods

Three ivorine teeth were prepared for all-ceramic crowns with finish lines placed equigingivally, 0.5 mm subgingivally, and 1.0 mm subgingivally. A single-cord technique was used for gingival retraction and a total of 180 intraoral scans were made using two intraoral scanners (Emerald; Planmeca USA Inc., Hoffman Estates, IL & Trios3; 3Shape A/S, Copenhagen, Denmark). The prepared teeth were separated from the dentoform and scanned using the same intraoral scanners to create reference scans. All scans were imported to the design software (Dental System 2019; 3Shape A/S, Copenhagen, Denmark). After marking the finish lines of prepared teeth, intraoral scans were aligned to the reference scans for comparisons. Vertical and horizontal marginal discrepancies were measured at 4 different measuring points (buccal, lingual, mesial, and distal) and analyzed. Two-way ANOVA and Tukey HSD tests were used for statistical analysis (α = 0.05).

Results

The average vertical and horizontal discrepancies from various groups ranged from -33μm to 440μm. For both intraoral scanners, subgingival finish line groups showed greater vertical and horizontal discrepancies compared with equigingival finish line groups. Saliva contamination significantly increased both vertical and horizontal discrepancies for all finish line locations. The discrepancy increases due to saliva contamination were greater for the subgingival groups.

Conclusions

Subgingival finish lines were not accurately captured using the intraoral scanners. The presence of saliva significantly reduced scanning trueness and this was amplified when the finish lines were located subgingivally.

This article is protected by copyright. All rights reserved

View on Web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Δημοφιλείς αναρτήσεις